Pressure Systems, Inc. Model 9046 User’s Manual

Page 13

www.PressureSystems.com

5.

Set the subnet mask and IP address if they are not correct:

● IP Address - use the address called for by the system.

● Set the subnet mask to 255.255.255.0

● Verify that the TCP Connect Port (Listen Port) is set to 9000.

6.

Click on the Save button. The window at the bottom of the Network Options Dialog Box
will indicate whether the operation was successful. If the information is not received, check
to see that the address was entered correctly.

Note

Clicking "Save" puts the new addresses in the module’s EEPROM,
but they will not be active until the module is re-booted or the power
is removed and re-applied.

7.

Click the Exit button. Right-click on the module icon, and select Disconnect.

8.

Remove power from the Model 9046, wait 5 seconds and re-apply.

Note

The computer and the module now will have incompatible IP
addresses. You will have to change the IP address to one that is
compatible. Exit the NUSS program, and change the computer IP.
Re-start NUSS. Observe the IP address when it is presented and
verify it.

9.

This completes the addressing procedure. Label the module with the new IP address.