Pressure Systems, Inc. Model 9046 User’s Manual

Page 58

www.PressureSystems.com

Note

Selecting too many data groups will reduce module performance.

bbbb

(hex) data selected for inclusion in each stream
packet

0001

(reserved for Valve Status Prefix (9046 does not use)

0002 Enable

UTR Alarm Status Prefix

(see alarm bit map

below)

0010 Enable

Primary Transducer (Temperature/Resistance)

EU Data

(default if "05" never executed after "00")

0020 Enable

Primary Transducer (Temperature/Resistance)

A/D Counts

0040 Enable

Primary Transducer (Temperature/Resistance)

Voltages

0080 Enable

UTR Temperature/Other EU Data (ûC)

0100 Enable

UTR Temperature/Other A/D Counts

0200 Enable

UTR Temperature/Other Voltages

Any UTR Alarm Status Prefix datum is delivered as a two-byte binary bit map (16-bit, big
endian) with each bit representing the status of UTR channel #16 through UTR channel #1
respectively. A bit value of 0 (zero) indicates the UTR is operating within its specified limits. A
value of 1 (one) indicates the UTR is outside its specified limits (i.e., "in alarm"). To be "in alarm"
the channel’s UTR datum value has to deviate from the average of the UTR data values of all
the TC channels, by an amount greater than a specified set point (in ыC).