Pressure Systems, Inc. Model 9046 User’s Manual

Page 34

www.PressureSystems.com

5.3 Reset (Command ’B’)

Purpose

: Instructs the module to reset internal operating parameters, and to set all internal

control variables to their default "reset" state (see description below). The current TCP/IP socket
connection will remain open. Execution after a power off/on cycle is optional (unnecessary).

Command

"B"
’B’ is the operation code.

Response

"A" or "Nxx"
’A’ for acknowledge; ’N’ for negative acknowledge,
error code xx.

Description

: The module returns to the following "reset" states if this command is executed:

• The Offset calibration adjustment coefficient for each transducer is set to the last value

stored in flash memory.

• The Gain calibration adjustment coefficient for each transducer is set to the last value

stored in flash memory.

• Number of Samples for Data Averaging is set to last value stored in flash memory

(factory default = 64).

• Any autonomous host data delivery streams defined by ’c’ sub-commands are reset

(undefined).

Example

:

Command module to reset "power-up" defaults:

"B"

Read following response:

"A"

Note

NUSS does not use this command normally. You may execute it
manually via the ’Execute Commands’ or ’Execute Script’ functions
of NUSS - though this should be avoided if NUSS is currently
operating the module via other NUSS functions.