Pressure Systems 9046 User Manual

Page 28

Advertising
background image

Pressure Systems, Inc. Model 9046 User’s Manual

Page 22

www.PressureSystems.com

Examples:

Correct the offset in channel 1 with channel 1 at 100 degree C :

Command:

h0001 100.0


Response:

0.294167 (offset coefficient returned)

Manually set channel 1offset coefficient to 0.4:

Command:

v00100 0.4

Normal
:

A


Response:

Error:

N08

Read channel 1 offset coefficient:

Command: u00100

Normal
:

0.4 (offset coefficient = 0.4)


Response:

Error:

N08

Store coefficients to FLASH PROM

Command:

w08


Response:

A