5 set sensor type – Pressure Systems 9046 User Manual

Page 20

Advertising
background image

Pressure Systems, Inc. Model 9046 User’s Manual

Page 14

www.PressureSystems.com

2.5 Set Sensor Type

1.

Confirm that the correct SCM is installed and jumpered correctly:

a. For Non-Universal SCMs:

Confirm that each of the 16 hardware channels contains the correct SCM for the
intended application. For isolated thermocouple applications, a SCM is required.

b. For Universal SCMs:

Confirm that each of the 16 SCMs is configured for the intended application. For thermocouple
applications, a jumper must be installed in the small sockets on the top of the SCM. (See Figure
1-2).

Signal Conditioning Module (SCM).) For all other applications, no jumper is required.

2.

Using Table 1 (on the next page), determine the value for each desired sensor type.

3.

Using NUSS configure each channel using a command of the format: vxxx51 yy, where
xxx

ranges from 001 through 010 HEX for channels 1 through 16, yy is the sensor type

value determined from Table 1.

Examples:

v00a51 18 will set channel 10 to a type T thermocouple.

V00951 60 will set channel 9 to an RTD with an ANSI 385 curve.

4.

Send a w07 command to write these values to non-volatile ROM.

Note

When the ‘v’ command is issued to set the sensor type for a
channel, the zero and offset coefficients for the channel are re-
initialized to the default values of 0.0 and 1.0 respectively.

Advertising