Hansol Multitech Uromat 3000 User Manual

Page 37

Advertising
background image

06/00 - 37 of 90 - 0116 7201

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

Install a spindle cross piece (Pos. 5+6) on each end of both support struts.
Remove the mounting screws (Fig. 07/Pos. 1+ Fig.06/Pos. 8) of the unit column from the
pallet and press it down onto the pallet.
Use the 4 lift spindles (Fig. 06/Pos. 7) to evenly raise the unit until the transport pallet
(Pos.9) is free and then slide it out from under the unit.
Install the 4 castors (Fig. 08/Pos. 3+4) (transport castors) for transport on the support studs
(Pos. 1+2). Use the 4 lift spindles to lower the unit (with narrow transport pathways, remove
the spindle cross pieces, Pos. 5+6) and move the unit to its installation location.
At the installation location, use the spindles to raise the unit until the castors are off the floor.
Remove the castors and evenly lower the unit.

Caution:
The two support struts on the installation frame (Fig. 06/Pos. 1+2) may be removed
only when the unit is secured to the floor.



Fig. 07

Fig. 08




2.4 Installing the Unit Column

Remove the lower transport bracket (Fig. 05/Pos. 2) on the tube unit support arm.
While observing the tensile strength data that is listed on them, use the 6 heavy-duty load
expansion bolts (for example, Liebig S12/40 or S12/65 expansion bolts) (Fig. 07/Pos. 1 and
Fig. 09/Pos. 2).
Remove the transport frame (Fig. 06/Pos. 1+2).
Disengage the tube unit support arm (Fig. 03/Pos. 3) towards the top out of the unit base
and carefully place it down on the floor next to the unit.
Disconnect the plug-in connector for the display.
Remove the cover panel on the unit base (Fig. 02/Pos. 1).