Technical maintenance – Hansol Multitech Uromat 3000 User Manual

Page 67

Advertising
background image

06/00 - 67 of 90 - 0116 7201

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

4. Technical Maintenance


NOTE:
The maintenance described below must be performed every12 months.
If unit components need to be moved using the motor during maintenance, or if functions
need to be checked while the unit is under power, the unit must be switched off immediately
after reaching the new position or after completing the particular work step and the
immediately disconnected from power again.
Components that are removed may only be replaced with original parts that are listed in the
attached spare parts list. For maintenance, only acid-free grease may be used. Do not
grease or oil bearings with seals.

4.1 Mechanical and Electrical Checks


Preparations:

- Move table to the max. up position.

- Switch power off to the system (line power)

- Remove the unit base cover cap
- Remove the unit base cover panels
- Remove the system carriage cover panels


Floor mounting:

- Check the unit column in both directions with the spirit level. Compensate any
differences by using shims and retighten the mounting screws.
- Check the mounting screws for secure tightness. If needed, retighten.
Torque 25 Nm (2.5 kpm)


Manual control unit:

- Check cable lead-in for good condition and for secure connection.
- Check the function of all buttons
- Check the function of all LED’s.


Multi-function footswitch:

- Check the cables for good condition and for secure connection.

If necessary, replace damaged cables.

- Clean the spaces between the individual footswitches.
- Check all switch functions.


Fluoroscopy and exposure footswitch:

- Check the cables for good condition and for secure connection.

If necessary, replace damaged cables.

- Clean the spaces between the individual footswitches.

- Check the switch functions.