06/00 - 57 of 90 - 0116 7201

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

Fig.71 Fig.72



3.19 Accessory Sensor Switches

Securely and audibly insert the table extension or the micturation seat into the
opening.
The corresponding microswitch is for sensing of the S17 table extension (Fig. 71/
Pos. 1) and S18 is for the micturation seat (Pos. 2). The switch strike plate must be
adjusted so that it is on the tip of the nib so that any play will trigger a positive
response of the S16 microswitch (Pos. 3).



3.20 Tabletop longitudinal Belt Tension


Belt tension should be determined as follows:
The spring balance should be placed centered between the belt return wheel (Fig. 72/Pos. 2)
and the belt tension roller (Pos. 3) to achieve a deflection of 1 cm when a force of 35 N is
applied. Adjustment can be made by means of the tension screw (Pos. 4) after loosening the
two mounting screws (Pos. 5+6).


3.21 Raise-Lower Tabletop Belt Tension

Belt tension should be determined as follows:
The spring balance should be placed centered between the drive wheel (on motor side) and
the belt wheel for the spindle (Fig. 73/Pos. 2) to achieve a deflection of 1 cm when a force of
80 N is applied. Adjustment can be made by means of the tension screw (Pos. 4) after
loosening the two mounting screws (Pos. 5+6).

Advertising