Hansol Multitech Uromat 3000 User Manual

Page 39

Advertising
background image

06/00 - 39 of 90 - 0116 7201

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

2.6 Temporarily Connecting Line Power


Caution:
If the unit is connected to a different line power than that listed on the order, the setting must
be changed to the one that corresponds to the local line power as specified by the Wiring
Diagram
to the power line transformer (Fig. 12/Pos. 1).

Provide the unit with power by connecting a temporary connection cable (plug-in cable) to
move the table for individual installation steps.
Connect the multi-function footswitch on the bottom of the table.





Fig. 13

Fig. 14



2.7 Installing the Tube Unit Support Arm

Tilt the tabletop all the way down into the horizontal position and move it towards the front,
transversely to the user, until it is in the end position. Move the tube unit carriage (system
carriage) max. to the head end.
Remove the drain panel (Fig. 13/Pos. 1) after loosening the 5 mounting screws (Pos. 2,3+4).
Place the tube unit support arm (Fig. 14/Pos. 1) on the longitudinal carriage (Pos. 2) and
secure it in place with the 6 screws (Pos. 3+4).
Remove the red transport bracket (Fig. 05/Pos. 1) (hooking device).
Reinstall the drain panel (Fig. 13/Pos. 1).









Advertising