Hansol Multitech Uromat 3000 User Manual

Page 69

Advertising
background image

06/00 - 69 of 90 - 0116 7201

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

- Check automatic stop in foot-end limit position
- Clean the guide rails.
- Check for play in the bearings. If necessary, readjust.


Tube unit support arm, exposure position, park position:

- Press the buttons on the control handle and move through the entire movement

range.

- Check for even movement during the unit movements.
- Check for engagement into the end positions.
- Clean the guide rails.
- Check for play in the bearings. If necessary, readjust them.
- Check for fluoroscopy and exposure blocking when the tube unit support arm is

not in the exposure position.



Component Panel:

- Switch power off to the system.
- Check all cables laid for good condition and secure connection.

If necessary, replace any damaged cables.

- Retighten any cable connections.
- Retighten the mounting screws of all components.
- Check P/C boards for secure seating.
- Check strain-reliefs and ground connections for secure connection.


Electronics:

- Electronic adjustment parameters per section.3.28


Accessories:

- Check all accessories present at the customer’s location.
- Hook in the table extension, tilt the table and check the factory set safety

distance of 12 cm; if necessary, adjust it.

- Check engagement of the leg support on the profile rail.
- Check for secure mounting of the head cushion mount.
- Check the patient table pad for good condition.
- Check the flush bowl for damage and check the function.


General:
- Reinstall the unit base cover panel.
- Reinstall the unit base cover panels.
- Reinstall the base cover panels.
- Reinstall the system carriage cover panels.
- Clean the unit.

- Clean surfaces with one of the following tested disinfecting agents:

Tego 103, Kosolin, Misty Multi-Purpose Disinfectant Cleaner, Misty Multi-Purpose
Disinfectant Cleaner II, Misty Disinfectant and Deodorant RTU, Precise Hospital
Foam Cleaner Disinfectant.
- Switch power back on.
- Carry out a function check according to 4.2.

Advertising