0116 7201 -84 of 90 - 06/00

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

5. Troubleshooting

Error

Number

Cause

Corrective Measure

F 1

Frequency converter 0 for tilt/vertical,
system failed.

Check connection cables, settings
(address, Baud rate).

F 2

Frequency converter 1 for tilt/vertical, table
longitudinal movement, cassette failed.

Check connection cables, settings
(address, Baud rate).

F 3

Frequency converters 0+1 failed.

Check connection cables, settings
(address, Baud rate).

F 10

Footswitch communication problem.

Check fuse, cabling.

F 11

Manual control unit communication
problem.

Check fuse, cabling.

F 13

Incorrect footswitch software version.

Replace footswitch.

F 14

Incorrect manual control unit software
version.

Replace manual control unit.

F 20

Analog voltage failed.

Check fuse.

F 21

M1 tilt/vertical potentiometer error.

Check potentiometer cables and
potentiometer.

F 22

System potentiometer error.

Check potentiometer cables and
potentiometer.

F 23

M3 tilt/vertical potentiometer error.

Check potentiometer cables and
potentiometer.

F 24

Table longitudinal potentiometer error.

Check potentiometer cables and
potentiometer.

F 25

Cassette potentiometer error.

Check potentiometer cables and
potentiometer.

F26

Table transverse potentiometer error /
motor defective.

Check motor potentiometer cables
and potentiometer.

F 28

Collimator vertical potentiometer error.

Check potentiometer cables and
potentiometer.

F 30

Button fuse malfunction.

Check buttons, cables.

F 31

Footswitch button sticking.

Check buttons.

F 32

Manual control unit button sticking.

Check buttons.

F 33

Exposure switch active during system start. Check switches.

F 34

Fluoroscopy switch active during system
start.

Check switches.

F 35

Exposure centering switch defective.

Check switches, cabling.