3.10 Tabletop Longitudinal Safety Switch

Lift up the tabletop (it will be held in place by the gas spring) and remove the cover
panel (Fig. 63/Pos. 1). Set the software end positions to 240

+/- 5

mm in each

direction. Adjust the limit switch (Fig. 64/Pos. 2) so that it switches off approx. 3 - 5
mm behind each end position. To do this, either the limit switch or the switch strike
place can be adjusted. Check the function over the entire movement path and
reinstall the cover panel.


3.11 Tabletop Transverse Safety Switch not configured