Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

F 40

Format switch cassette incorrect.

Check 4 format switches.

F 41

M1 motor breaker stuck/defective.

Check K1 breaker.

F 42

M2 motor breaker stuck /defective.

Check K2 breaker.

F 43

M3 motor breaker stuck /defective.

Check K3 breaker.

F 44

M4 motor breaker stuck /defective.

Check K4 breaker.

F 45

M5 motor breaker stuck /defective.

Check K5 breaker.

F 46

M6 motor relay stuck /defective.

Check K19 relay (CPU board).

F 47

FU0 N>0 active.

FU0 relay defective/ input
defective / check cabling.

F 48

FU1 N>0 active.

FU1 relay defective / input
defective / check cabling.

F 50

Collimator malfunction.

Check cabling, potentiometers,
setup parameters.

F 61

Limit switch, temperature, breaker
malfunction, M1 tilt /vertical drive or M3.

Free run (press button on breaker
board). Allow motor to cool down.

F 62

Limit switch, temperature, breaker
malfunction, M2 system drive.

Free run (press button on breaker
board). Allow motor to cool down.

F 64

Limit switch, temperature, breaker
malfunction, M4 table longitudinal drive.

Free run (press button on breaker
board). Allow motor to cool down.

F 79

Invalid parameter version

Initialize correct parameter
version.

F 80

EEPROM memory error.

Download new parameter file.

F 99

Unit is in the service mode.

Set the service switch to the RUN
mode.



F = Fatal Error = System error = cannot be corrected