Replacement parts – Fluke 45 User Manual

Page 101

Advertising
background image

Maintenance

Replacement Parts

6

6-7

Table 6-3. Performance Tests for mA Current Functions

Display

Function

Range (Rate)

Input Level

Frequency

Min

Max

[

(mA)

\

(mA)

30 mA

100 mA

30 mA

100 mA

+30 mA

+100 mA

30 mA

100 mA

1 kHz

1 kHz

29.982

99.93

29.840

99.40

30.018

100.07

30.160

100.60

Table 6-4. Performance Tests for A Current Functions

Display

Function

Range (Rate)

Input Level

Frequency

Min

Max

[
\

10 A

10 A

+10 A

10 A

1 kHz

9.975

9.890

10.025

10.110

Replacement Parts

Parts that can be replaced by the user are listed in Table 6-5. Other parts must be replaced
by qualified service personnel only, following the procedures in the Service Manual.

Replacement parts and price information can be obtained from Fluke or an authorized
Fluke distributor. To ensure prompt and efficient handling, include the following
information with your order:

Quantity

Fluke Part Number

Part Description

Instrument Model and Serial number.

Table 6-5. Replacement Parts

ITEM

DESCRIPTION

FLUKE PART NO.

C40*

F1

W

F2

W

F3

W

F5

W

RS40*

Soft Carrying Case (Optional)

mA Input Fuse. F 500 mA, 250 V, (Fast blow), 1500 A breaking
capacity

10 A Input Fuse. F 15 A, 250 V, (Fast blow), 10,000 A minimum
breaking capacity; Or, F 15 A, 600 V, (Fast blow), 10,000 A minimum
breaking capacity 820829

Line Fuse. T 125 mA, 250 V, (Slow blow)

See Supplement

RS-232 terminal interface cable. Connects Fluke 45 to any properly
configured DTE connector (DB-25, male pins). IBM PC or IBM
PC/XT, IBM PS/2 (Models 25,30,50, 60,70,80).

838151

830802

822254

Advertising