Fluke 45 User Manual

Page 34

Advertising
background image

Operating the Meter From the Front Panel

Ranging

3

3-7

Table 3-3. Current Ranges and Full Scale Values

Fast Reading Rate

Medium Reading Rate

Slow Reading Rate

Range

Full Scale

Range

Full Scale

Range

Full Scale

30 mA

30.00 mA

30 mA

30.000 mA

10 mA

9.9999 mA

100 mA

100.0 mA

100 mA

100.00 mA

100 mA

99.999 mA

10 A

10.00 A*

10 A

10.000 A*

10 A

9.9999 A

* 20 A for maximum of 30 seconds

Table 3-4. Ohms Ranges and Full Scale Values

Fast Reading Rate

Medium Reading Rate

Slow Reading Rate

Range

Full Scale

Range

Full Scale

Range

Full Scale*

300

300.0

300

300.00

100

98.000

3 k

3.000 k

3 k

3.0000 k

1000

980.00

30 k

30.00 k

30 k

30.000 k

10 k

9.8000 k

300 k

300.0 k

300 k

300.00k

100 k

98.000 k

3 M

3.000 M

3 M

3.0000 M

1000 k

980.00 k

30 M

30.00 M

30 M

30.000 M

10 M

9.8000 M

300 M

300 M

300 M

300.0 M

100 M

98.0 M

**

*Typical
** Because of the method used to measure resistance, the 100 M (slow) and 300 M (medium and fast)
ranges cannot measure below 3.125 M and 20 M respectively. "UL" (Underload) is shown on the display
for resistances below these nominal points, and the computer interface outputs "+1 E-9".

Table 3-5. Frequency Ranges and Full Scale Values (Slow and Medium*)

Range

Full Scale

1000 Hz

999.99 Hz

10 kHz

9.9999 kHz

100 kHz

99 999 kHz

1000 kHz

999.99 kHz

1 MHz

9.9999 MHz

* Fast (F) reading rate has one digit of resolution less.