Fluke 45 User Manual

Page 69

Advertising
background image

45

Users Manual

5-10

In the following examples, <space> between commands is for readability only. However a
<space> between a command and its argument is required. The PRINT command sends a
terminator at the end of the string.

PRINT @3, "*RST; VAC; FREQ 2; RANGE 4; TRIGGER 2; *TRG; VAL1?"

IEEE-488 bus address

Reset to power-up configuration

Select frequency
for secondary display

Select volts ac function
for primary display

Select external trigger
(Rear panel trigger disabled, no trigger delay)

Trigger a reading

Return reading shown on primary display

Select 300V range for primary display

The following string configures the meter and triggers a volts ac reading in primary
display and a frequency reading in the secondary display.

EXAMPLE 2:

The following string configures the meter and triggers an ohms reading to be shown
on the primary display.

EXAMPLE 1:

PRINT @3, "*RST; OHMS; RANGE 1; RATE M; TRIGGER 2; *TRG; VAL1?"

IEEE-488 bus address

Reset to power-up configuration

Select 300 range

Select function

Select external trigger
(Rear panel trigger disabled, no trigger delay)

Trigger a reading

Return reading shown on primary display

Select medium reading rate

W

W

aam19f.eps

Figure 5-1. Typical IEEE-488 Input Strings