Neither function modifiers (REL, dB, HOLD, and MN MX) nor the manual range mode
can be selected in the secondary display. Measurement ranges in the secondary display
are always selected through autoranging.

Input Terminals

The input terminals, shown in Figure 3-4, are located on the left of the front panel.

The meter is protected against overloads up to the limits shown in Table 3-1. Exceeding
these limits poses a hazard to both the meter and operator.

CAL

ENABLE

V

10A

100

mA

600V CAT I

1000V CAT I

FUSE F1

500 mA

F 250V

POWER

AUTO

2ND

MN MX

RATE

dB

REL

REF#

LOCAL

THRESH

ADDR

BAUD

HOLD

LO

HI

COMP

A

A

V

FREQ

DUAL DISPLAY MULTIMETER

45

V

!

REF

FUSED

COM

V

10A

100

mA

600V CAT I

1000V CAT I

FUSE F1

500 mA

F 250V

!

FUSED

COM

Volts, Ohms, Diode Test Input Terminal

Amperes Input Terminal. For Current Measurements up to
10A continuous (or 20A for 30 sec).

Milliamperes Input Terminal. For Current Measurements
up to 100mA.

Common Terminal. Return Terminal for all Measurements.

Note: Frequency Measurements are Normally taken

from the V Input, although the 10A and
100mA Inputs can be used.

aam12f.eps

Figure 3-4. Input Terminals