Fluke 45 User Manual

Page 50

Advertising
background image

Applications

Using the Dual Display

4

4-3

Table 4-1. Sample Dual Display Applications

Primary Display

Secondary Display

Applications

Volts DC

Volts AC

Monitor dc level and ac ripple of power supply
Troubleshoot amplifier circuits

Volts DC

Current DC

Check power supply load regulation

Monitor UUT current draw and circuit voltages

Monitor loop current and voltage drop across
transmitter

Volts DC

Current AC

Line and load regulation tests

dc/ac or ac/dc converters

Volts AC

Current DC

Line and load regulation tests

dc/ac or ac/dc converters

Volts AC

Current AC

Line and load regulation tests

Transformer (magnetic circuit) saturation

Volts AC

Frequency

Measure ac amplitude and frequency for line
voltage and ac signal analysis

Measure frequency response of an amplifier

Adjust ac motor control

Read noise in telecommunication applications

Adjust portable power generator to optimize
power output

Set frequency compensation for a network

dB (in Volts dc)

Frequency

Use in "print-only" mode (see "RS-232 Print-
Only Mode" in Chapter 5) for quick Bode plots
(frequency vs. amplitude)

Test frequency response

Current DC

Current AC

Measure ripple and dc current draw of switching
power supply

Measure current dissipation in protective fuse
resistors used in power supplies

Measure ripple and noise on a line

MN MX

Actual Value

Show the minimum or maximum value recorded
and the present measurement

REL

Actual Value

Show actual measurement and the difference
between this value and the relative base.

REL

Resistance

Select and sort resistors. (See also "Using the
Compare Function" in Chapter 3.)

HOLD

Actual Value

Show actual measurement while holding a
previous, stable measurement on the primary
display

Advertising