Replacing the 100 ma input fuse, Replacing the 10 a input fuse – Fluke 45 User Manual

Page 97

Advertising
background image

Maintenance

Current Input Fuses

6

6-3

If the fuse is good, the meter will read between .04

and 1.0

. If the fuse is blown,

the meter will read >10 M

to OL.

Replacing the 100 mA Input Fuse

W

Warning

For protection against fire or arc flash, replace a blown fuse
only with one of identical rating.

The 100 mA input fuse is mounted in the front panel input jack (see Figure 6-2).

To replace this fuse, first unplug the line cord. Then press in on the input jack and turn it 90
degrees counter-clockwise as shown in Figure 6-2. Slide out the fuse holder and fuse.

Replace a blown fuse with one of identical rating (see Table 6-5) and reinsert the fuse and
holder into the input terminal socket. Secure the fuse holder by pressing in and turning the
holder 90 degrees clockwise.

1000V CAT

FUSE F1

500 mA
F 250V

FUSED

600V C

AT

10A

100

mA

V

COM

Front Panel Input Terminal

100 mA Input Socket

Fuse Holder

To remove, push in and turn counter clockwise.
To insert, reverse this procedure.

F1 Fuse (500 mA, 250V, Fast Blow)
1500 A Minimum Breaking Capacity

qb13f.eps

Figure 6-2. Replacing the 100 mA Input Fuse

Replacing the 10 A Input Fuse

The 10 A input fuse is located inside the meter. To replace the fuse:

1.

Remove the single Phillips-head screw on the bottom of the case and the Phillips -
head screw on each side of the rear bezel.

W

Warning

Opening the case may expose hazardous voltages. To avoid
electric shock, always disconnect the power cord and
measuring inputs before opening the case.

2.

Remove the bezel and slip the case back from the front of the meter. The fuse and
fuse clip are visible at the front of the main printed circuit assembly (pca) near the
input terminals.