Fluke 45 User Manual

Page 110

Advertising
background image

45

Users Manual

Accuracy

Accuracy

Range

Frequency

Slow

Medium

Fast

mA (To 100 mA)

mA (To 100 mA)

mA (To 100 mA)

A (1-10A)

A (1-10A)

A (0.5 to 1A)

A (0.5 to 1A)

20-50 Hz

50 Hz-10 kHz

10 -20 kHz

20-50 Hz

50 Hz-2 kHz

20-50 Hz

50Hz-2 kHz

2 % + 100

0.5 % + 100

2 % + 200

2 % + 100

1 % + 100

2 % + 300

1 % + 300

2 % + 10

0.5 % + 10

2 % + 20

2 % + 10

1 % + 10

2 % + 30

1 % + 30

7 % + 2

0.8 % + 2

2 % + 3

7 % + 2

1.3 % + 2

7 % + 4

1.3 % + 4

mA accuracy specifications apply within the following limits, based on reading rate:

Slow Reading Rate:

Between 15,000 and 99,999 counts (full range)

Medium Reading Rate:

Between 1,500 and 30,000 counts (full range)

Fast Reading Rate:

Between 150 and 3,000 counts (full range)

Maximum Crest Factor

3.0

Maximum Input

To be used in protected, low energy circuits only, not to exceed 250 V or 4800
Volt-Amps. (IEC 664 Installation Category II.)

mA

300 mA dc or ac rms. Protected with a 500 mA, 250 V, IEC 127-sheet 1, fast
blow fuse

A

10 A dc or ac rms continuous, or 20 A dc or ac rms for 30 seconds maximum.
Protected with a 15 A, 250 V, 10,000 A interrupt rating, fast blow fuse.

Note

Resistance between the COM binding post and the meter’s internal
measuring circuits is approximately .003

.