Frequency – Fluke 45 User Manual

Page 112

Advertising
background image

45

Users Manual

Open Circuit Voltage

3.2 V maximum

Continuity Capture Time

50 us maximum, 10 us typical

Input Protection

500 volts dc or rms ac

Note

When the meter is set to measure frequency and there is no input signal
(i.e., input terminals are open), the meter may read approximately 25 kHz
(rather than the expected zero). This is due to internal capacitive pickup of
the inverter power supply into the high-impedance, input circuitry. With
source impedance of <2 k

, this pickup will not affect the accuracy or

stability of the frequency a reading.

Frequency

Frequency Range

5 Hz to >1 MHz

Applicable Functions

Volts ac and Current AC

Resolution

Range

Slow & Medium

Fast

Accuracy

1000 Hz

10 kHz

100 kHz

1000 kHz

1 MHz*

.01 Hz

.1 Hz

1 Hz

10 Hz

100 Hz

.1 Hz

1 Hz

10 Hz

100 Hz

1 kHz

05% + 2

.05% + 1

.05% + 1

.05% + 1

Not Specified

* Specified to 1 MHz, but will measure above 1 MHz.

Sensitivity of AC Voltage

Frequency

Level (sine wave)

5 Hz-100 kHz

100 kHz - 300 kHz

300 kHz - 1 MHz

Above 1 MHz

30 mV rms

100 mV rms

1 V V rms

Not specified

Sensitivity Level of AC Current

Frequency

Input

Level

5 Hz-20 kHz

45 Hz-2 kHz

100 mA

10 A

>3 mA rms

>3 A rms