Selecting a measurement function, Ranging, Autoranging – Fluke 45 User Manual

Page 32: Ranging ( e , u , t )

Advertising
background image

Operating the Meter From the Front Panel

Selecting a Measurement Function

3

3-5

Table 3-1. Input Limits

FUNCTION

INPUT TERMINALS

MAXIMUM INPUT

|

V

eG

and COM

1000V dc

{

and FREQ

V

eG

and COM

750V ac rms, 1000V peak, 2 x 10

7

V-Hz

normal mode, or 1 x 10

6

V-Hz common

mode (whichever is less)

mA

D

and FREQ

100 mA and COM

300 mA dc or ac rms

\

and FREQ

10A and COM

10A dc or ac rms (or 20A dc or ac rms for
30 sec. Max)

e

V

eG

and COM

500V dc or ac rms on all ranges

G

V

eG

and COM

500V dc or ac rms

All Functions

Any terminal to earth

1000V dc or peak ac

Selecting a Measurement Function

Press a function button (white), shown in Figure 3-5, to select a measurement function To
select ac + dc total rms readings, press

Z and Y, or D and C, simultaneously.

When you select a function, annunciators turn on to indicate the function selected. If a
reading is shown on the secondary display when a function button is pressed, the
secondary display will be turned off.

Ranges and full scale values are summarized in Table 3-2 for voltage, Table 3-3 for
current, Table 3-4 for ohms, and Table 3-5 for frequency.

Ranging

(

E, U, T)

Ranging operations are performed using the

E, U, and T buttons (see Figure 3-6).

Autoranging

When you are in the autorange mode, the AUTO annunciator is lit.

In autorange, the meter automatically selects the next higher range when a reading is
greater than full scale. If no higher range is available, "OL" (overload) is displayed on the
primary or secondary display. The meter automatically selects a lower range when a read-
ing is less than approximately 9 % of full scale.