Dc voltage – Fluke 45 User Manual

Page 106

Advertising
background image

45

Users Manual

DC Voltage

Resolution

Accuracy

Range

Slow

Medium

Fast

(6 Months)

(1 Year)

300 mV

3 V

30 V

300 V

1000 V

10 µV

100

µ

V

1

mV

10

mV

100 mV

100 µV

1

mV

10 mV

100 mV

1 V

002 % + 2

0.02 % + 2

0.02 % + 2

0.02 % + 2

0.02 % + 2

0.025 % + 2

0.025 % + 2

0.025 % + 2

0.025 % + 2

0.025 % + 2

100 mV

1000 mV

10 V

100 V

1000 V

1 µV

10 µV

100

µ

V

1 mV

10 mV

0.02 % + 6

0.02 % + 6

0.02 % + 6

0.02 % + 6

0.02 % + 6

0.025 % + 6

0.025 % + 6

0.025 % + 6

0.025 % + 6

0.025 % + 6

Input Impedance

10 M

e in parallel with <100 pF

Note

In the dual display mode, when the volts ac and volts dc functions are
selected, the 10 M

dc input divider is in parallel with the 1 M

ac

divider.

Normal Mode Rejection Ratio

>80 dB at 50 Hz or 60 Hz, slow and medium rates

>54 dB for frequencies between 50-440 Hz, slow and medium rates

>60 dB at 50 Hz, fast rate (Note: Fast rate has no filtering)

Maximum Allowable AC Voltage While Measuring DC Voltage or (AC + DC)
Voltages

Peak Normal Mode Signal

Range

Max Allowable Peak AC

Voltage

NMRR* >80 dB†

NMRR >60 dB†

300 mV

3 V

30 V

300 V

1000 V

100 mV

1000 mV

10 V

100 V

1000 V

15 V

15 V

1000 V

1000 V

1000 V

15 V

15 V

50 V

50 V

200 V

15 V

15 V

300 V

300 V

1000 V

* NMRR is the Normal Mode Rejection Ratio
† Normal Mode Rejection Ratio at 50 Hz or 60 Hz

±

0.1 %

Common Mode Rejection Ratio

>90 dB at do, 50 or 60 Hz, (1 k

e unbalanced, medium and slow rates)