Ohms, Diode test/continuity – Fluke 45 User Manual

Page 111

Advertising
background image

Appendices

Specifications

A

A-7

Ohms

Resolution

Range

Slow

Medium

Fast

Accuracy

Typical Full

Scale

Voltage

Max Current
Through the

Unknown

300

3 k

30 k

300 k

3 M

30 M

300 M

*

100

1000

10 k

100 k

1000 k

10 M

100 M

*

1 m

10 m

100 m

1

10

100

100 k

10 m

100 M

1

10

100

1 k

100 k

100 M

1

10

100

1 k

10 k

1 M

0.05 % + 2 + 0.02

0.05 % + 2

0.05 % + 2

0.05 % + 2

0.06 % + 2

0.25 % + 3

2 %

0.05 % + 8 + 0.02

0.05 % + 8 + 0.02

0.05 %+8

0.05 % + 8

0.06 % + 8

0.25 % + 6

2 % + 2

0.25

0.24

0.29

0.29

0.3

2.25

2.9

0.09

0.10

0.11

0.11

0.12

1.5

2.75

1 mA

120µA

14 µA

1.5 µA

150 µA

320 µA

320 µA

1 mA

120 µA

14 µA

1.5 µA

150 µA

150 µA

320 µA

*Because of the method used to measure resistance, the 100 M

(slow) and 300 M

(medium and fast)

ranges cannot measure below 3.2 M

and 20 M

, respectively. "UL" (underload) is shown on the display for

resistances below these nominal points, and the computer interface outputs "+1 E-9".

Open Circuit Voltage

3.2 V maximum on the 100

, 300

, 30 M

, 100 M

, and 300 M

ranges, 1.5 V

maximum on all other ranges.

Input Protection

500 V dc or rms ac on all ranges

Diode Test/Continuity

Maximum Reading

Resolution

Slow

Medium

Fast

999.99 mV

2.5 V

2.5 V

10 µV

100 µV

1 mV

Test Current

Approximately 0.7 mA when measuring a forward biased junction.

Audible Tone

Continuous tone for continuity. Brief tone for normal forward biased diode or
semiconductor junction.