Chapter 1 Introduction

Series 24-HP Instruction Manual

1-6

IM-24-HP

As shown in Figure 1-3, Innova-Mass™ meters exhibit a constant
Strouhal number across a large range of Reynolds numbers, indicating a
consistent linear output over a wide range of flows and fluid types. Be-
low this linear range, the intelligent electronics in Innova-Mass automat-
ically corrects for the variation in the Strouhal number with the Reynolds
number. The meter’s smart electronics corrects for this non-linearity via
its simultaneous measurements of the process fluid temperature and pres-
sure. This data is then used to calculate the Reynolds number in real
time. Innova-Mass meters automatically correct down to a Reynolds
number of 5,000.

0.3

0.2

0.1

0.0

3

4

10

Linear range

Reynolds Number, Re

Strouhal Number, St

10

5

10

10

8

10

6

10

7

5000

Corrected range

Figure 1-3. Reynolds Number Range for the Innova-Mass

Temperature Measurement

Innova-Mass Flow Meters use a 1000 ohm platinum resistance temperature
detector (PRTD) to measure fluid temperature.