Series 24-HP Instruction Manual

Chapter 2 Installation

IM-24-HP

2-17

Adjusting Meter Orientation

Depending on installation requirements, you may need to adjust the meter
orientation. There are two adjustments available. The first rotates the posi-
tion of the LCD display/keypad and is available on both in-line and inser-
tion meters. The second is to rotate the enclosure position. This adjust-
ment is only allowed on Series 240 In-Line meters.

Display/Keypad Adjustment (All Meters)

Figure 2-12. Display/Keypad Viewing Adjustment

The electronics boards are electrostatically sensitive. Wear a grounding
wrist strap and make sure to observe proper handling precautions required
for static-sensitive components. To adjust the display:

1. Disconnect power to the flow meter.

2. Loosen the small set screw which secures the electronics enclosure.

Unscrew and remove the cover.


3. Loosen the 4 captive screws.

4. Carefully pull the display/microprocessor board away from the meter

standoffs. Make sure not to damage the connected ribbon cable.


5. Rotate the display/microprocessor board to the desired position. Max-

imum turn, two positions left or two positions right (180-degrees).


6. Align the board with the captive screws. Check that the ribbon cable

is folded neatly behind the board with no twists or crimps.

7.

Tighten the screws. Replace the cover and set screw. Restore power to
the meter.