Chapter 3 Operation

Series 24-HP Instruction Manual

3-8

IM-24-HP

Totalizer #1 Menu

Password

ENTER

Use

keys to access menus

Totalizer

Menu

Totaling

Inactive

Mass

Volume

ENTER

Run Mode

(unit)/Pulse

xxxx

Reset Total?

YES or NO

Example:

Maximum flow rate = 600 gallons per minute

(600 gallons per minute = 10 gallons per second)

If unit per pulse is set to 600 gallons per pulse,

the totalizer will pulse once every minute.

If unit per pulse is set to 10 gallons per pulse,

the totalizer will pulse once every second.

Energy

Use the Totalizer Menu to configure and monitor the totalizer. The totalizer output is a 50 millisecond
(.05 second) positive pulse (relay closed for 50 milliseconds). The totalizer cannot operate faster than
one pulse every 100 millisecond (.1 second). A good rule to follow is to set the unit per pulse value
equal to the maximum flow in the same units per second. This will limit the pulse to no faster than
one pulse every second.

Example for Setting the Totalizer

The following shows how to set the totalizer to track mass flow in kg/sec. (All outputs are disabled while using
the Setup Menus.)

First, set the desired units of measurement:

1. Use ÕÖ keys to move to the Units Menu (see to page 3-9).
2. Press Ø key until Mass Flow Unit appears. Press

ENTER

.

3. Press Ø key until kg appears in the numerator. Press Ö key to move the underline cursor to the

denominator. Press the Ø key until sec appears in the denominator. Press

ENTER

to select.

4. Press × key until Units Menu appears.

Second, set the pulse output:

1. Use ÕÖ keys to move to the Totalizer Menu.
2. Press the Ø key until Totaling appears.
3. Press

ENTER

and press the Ø key to select Mass. Press

ENTER

.

4. Press Ø key to set the pulse output in the units you have selected for mass flow of kg/sec. Press

ENTER

and use ЧШХЦ keys to set the pulse value equal to the maximum flow in the same units per

second. Press

ENTER

.

5. To reset the totalizer, press Ø key until Reset Total? appears. Press

ENTER

and the Ø key to reset the

totalizer if desired. Press

ENTER.

6. Press the

EXIT

key and answer

YES

to permanently save your changes.