Example for Setting an Output
The following shows how to set Output 1 to measure mass flow with 4 mA = 0 lb/hr and 20 mA = 100 lb/hr with
a time constant of 5 seconds. (All outputs are disabled while using the Setup Menus.)
First, set the desired units of measurement:
1. Use ÕÖ keys to move to the Units Menu (see page 3-9).
2. Press Ø key until Mass Flow Unit appears. Press
ENTER
.
3. Press Ø key until lb appears in the numerator. Press Ö key to move the underline cursor to the
denominator. Press the Ø key until hr appears in the denominator. Press
ENTER
to select.
4. Press × key until Units Menu appears.
Second, set the analog output:
1. Use ÕÖ keys to move to the Output Menu.
2. Press the Ø key until 4-20 mA Output 1 appears.
3. Press Ö key to access Measure selections. Press
ENTER
and press the Ø key to select Mass. Press
ENTER
.
4. Press Ö key to set the 4 mA point in the units you have selected for mass of lb/hr. Press
ENTER
and
use ЧШХЦ keys to set 0 or 0.0. Press
ENTER
.
5. Press Ö key to set the 20 mA point. Press
ENTER
and use ЧШХЦ keys to set 100 or 100.0. Press
ENTER
.
6. Press Ö key to select the Time Constant. Press
ENTER
and use ЧШХЦ keys to select 5. Press
ENTER
.
7. Press the
EXIT
key and answer
YES
to permanently save your changes.