Ethernet power-on defaults, Installing the vxipc 800 series, Figure 3-14. ethernet power-on defaults – National Instruments 800 Series User Manual

Page 37

Advertising
background image

Chapter 3

VXIpc 800 Series Configuration and Installation

© National Instruments Corporation

3-13

VXIpc 800 Series User Manual

Ethernet Power-On Defaults

The VXIpc-800 Ethernet circuitry loads its power-on settings from an
EEPROM. You should not change switch S1 from its default setting;
however, the switch setting required to do so is included in Figure 3-14
for informational purposes.

Figure 3-14. Ethernet Power-On Defaults

Installing the VXIpc 800 Series

This section contains general installation instructions for the
VXIpc-800. Consult your VXIbus mainframe user manual or technical
reference manual for specific instructions and warnings.

1.

Plug in your mainframe before installing the VXIpc-800. The
power cord grounds the mainframe and protects it from electrical
damage while you are installing the module.

Warning: To protect both yourself and the mainframe from electrical hazards,

the mainframe should remain off until you are finished installing the
VXIpc-800 module.

2.

Remove or open any doors or covers blocking access to the
mainframe slots.

3.

If you are installing the VXIpc-800 into a D-size mainframe, install
a support designed for installing C-size boards in D-size
mainframes. The VXIpc-800 has no P3 connector and cannot
provide P3 Slot 0 control to VXI devices requiring this capability.

Caution:

If the VXIpc-800 is not configured for automatic Slot 0 detection, be
certain that the slot you select in your VXIbus mainframe matches the
VXIpc-800 configuration as either a Slot 0 device or a Non-Slot 0 device.
If you install your VXIpc-800 into a slot that does not correspond with the
jumper setting, you risk damage to the VXIpc-800, the VXIbus backplane,
or both.

S1

S1

a. Ethernet Power-On Defaults

Loaded from EEPROM

b. Hardwired Ethernet Values

!

Advertising