Table 7. ni 654x/655x/656x smb connectors, Table 8. active led indicators, Bles 8 – National Instruments PXI-6561/6562 User Manual

Page 25: Ble 7, and

Advertising
background image

© National Instruments Corporation

25

NI Digital Waveform Generator/Analyzer Guide

Tables 8 and 9 describe the different conditions indicated by the two
NI PXI LEDs.

1, 4, 7, 10, 13,
16, 19, 22, 25,
28, 31, 34, 37,
40, 43, 46, 49,
52, 55, 58

GND

Ground

Ground reference for signals.

11, 12, 68, 69

RESERVED

N/A

These terminals are reserved for
future use. Do not connect to
these pins.

Table 7. NI 654X/655X/656X SMB Connectors

Connector

Signal

Name

Signal

Type

Description

CLK IN

Reference/
Clock Input

Control

Terminal for the external Reference
clock used for the PLL or for the
external Sample clock used for
pattern generation and/or
acquisition.

PFI 0

Programmable
Function
Interface
(PFI) 0

Control

Input terminal to the device for
external triggers, or output terminal
from the device for events.

CLK OUT

Reference/
Clock Output

Control

Terminal for the exported PLL
Reference clock or the exported
Sample clock.

Table 8. ACTIVE LED Indicators

Color

Indications

Off

Device not armed, not triggered, or experiencing an error.

Amber

Device armed and awaiting Start trigger. If performing a
dynamic acquisition operation, the device may be acquiring
pretrigger samples.

Green

Device received Start trigger.

Red

Error condition.

Table 6. NI 656X DDC Connector Pins (Continued)

Pins

Signal Name

Signal

Type

Signal Description

Advertising