Section 4. Remote Operation
Readback and Status Indicators

50

Operating Manual for XPD Series Power Supply

Remote

Programming

Options

See

Table 4.1

for a summary of the options available to you for programming output

voltage and current limit using an analog source.

Table 4.1 Remote Programming Options

Programming

Output

Voltage with a

0-10 Vdc

Voltage

Source

1. Connect J210 pin 6 (remote voltage program select) to J210 pin 8 (auxiliary

ground).

2. Connect the control voltage source between pin 14 (voltage program) and pin 13

(program return).

3. Vary the external voltage from 0-10 Vdc to cause the power supply output to

vary from 0-100% of rated output voltage.

Programming

Output

Current with

a 0-10 Vdc

Voltage

Source

1. Connect J210 pin 7 (remote current program select) to J210 pin 8 (auxiliary

ground).

2. Connect the control voltage source between J210 pin 5 (current program,

positive) and pin 8 (program return).

3. Vary the external voltage from 0-10 Vdc to cause the power supply current limit

to vary from 0-100% of rated output.

Readback and Status Indicators

Voltage and

Current

Readback

Voltage Readback

Connect a voltage monitoring device between J210

connector pin 15 (voltage readback - positive) and pin 13 (program return -
negative).

Range: 0-10 V = 0-100% full rated voltage output of power supply

Current Readback

Connect a voltage monitoring device between J210

connector pin 12 (current readback - positive) and pin 13 (program return -
negative).

Range: 0-10 V = 0-100% full rated current output of power supply

Control of...

Programming Scale

Output voltage

0-10 V voltage sources

Current limit

0-10 V voltage sources