Overview

80

Electrodes

Socket acc. to DIN 19 262 or BNC
4-mm socket for reference electode

Temp. probe

4-mm socket for temperature probe

Interface/
analog output

Bidirectional RS 232 interface or ana-
log output with autom. recognition of
the PC, printer, or recorder connected

Plug-in power
supply (option)

2 pole special Friwo

Connectors

Input impedance

˜

10

12

W

Offset current

ˆ

10

-12

A

AutoCal TEC

With Technical buffers, pH values
2.00; 4.00; 7.00; 10.00 at 25 °C;
buffers are temperature compensa-
ted in the range 0 ... 90 °C

AutoCal DIN

With standard buffers according to
DIN 19 266/NIST, pH values:
1.68; 4.01; 6.87; 9.18 at 25 °C;
buffers are temperature compensa-
ted in the range 0 ... 90 °C

ConCal

With any buffers; one and two-point
calibration, manual acceptance of
measured values and temp. input

Input amplifier

Calibration modes