Operation

104

8 Thoroughly rinse the electrode with distilled water.

9 If necessary, set the temperature of the second buffer

solution by pressing <

> <

>.

10 Submerse the electrode in the second buffer solution

.

11 Press the <run/enter> key.

The second measured pH value appears on the dis-
play.

12 Set the nominal pH value of the second buffer solu-

tion (at the current temperature).

13 Press the <run/enter> key.

The value of the slope (mV/pH) appears on the dis-
play.
The sensor symbol shows the evaluation of the elec-
trode after the two-point calibration.

14 Press the <run/enter> key.

The value of the asymmetry (mV) appears on the dis-
play again.

15 To return to the measuring mode: Press the <pH>

key.