3 transmitting measured values – Xylem CG 843 User Manual
Page 24
Operation
92
4.2.3 Transmitting measured values
You can transmit measured values (data records) in 3 ways:
l
Switch on the data transmission (Int 2) (see 114)
– After expiry of the selected interval, the current data
record is sent to the RS interface.
l
Switch on AutoStore (Int 1) (see section 4.4.2 S
– After expiry of the selected interval, the current data
record is sent to the RS interface and in addition is
stored in the data store of the instrument
– AutoStore (Int 1) covers the
data transmission
interval
(Int 2).
l
Press the <run/enter> key
This manually triggers a transmitting of the current mea-
sured values at any time - independently of the selected
intervals.
Note
If you connect a recorder (analog output), both the output to
the digital output are switched off.