Operation

122

4.7

Reset

You can reset (initialize) measuring and configuration
parameters separately from one another.

Measuring parame-

ters

The following measuring parameters (pH InI) are reset to the
values they had on delivery:

Note
When the measuring parameters are reset, the calibration
data are lost. After the parameters have been reset, cali-
brate!

Configuration

parameters

The following configuration parameters (InI) are reset to the
values they had on delivery:

Measuring mode

pH

Asymmetry

0 mV

Slope

-59.16 mV/pH

Calibration procedure

AutoCal TEC

Temperature, manual

25 °C

Resolution of pH display

0.01

Baud rate

4800

Interval 1
(automatic storing)

OFF

Interval 2
(for data transmission)

OFF