4.3.1 Calibration interval (Int 3)

The flashing sensor symbol reminds you to calibrate regu-
larly. After the selected calibration interval (Int 3) expires,
the sensor symbol flashes. Measurements can continue.

Note
To ensure the high measuring precision of the measuring
system, perform a calibration after the calibration interval
expires.

Setting the calibra-

tion interval

The calibration interval (Int 3) is set to 7 days in the factory.
The interval can be changed (1 ... 999 days):

1 Switch off the instrument.

2 Press <pH> and hold down the key.

3 Press the <on/off> key.

The

display test

appears briefly on the display.

The instrument then switches automatically to the
configuration level.

4 Press the <run/enter> key until

Int 3

appears on the

display.

5 Press <

> <

> to set the required time interval until

the next calibration.

6 Confirm with <run/enter>.

7 Change to the measuring mode by pressing <pH>.