Operation
100
4.3.3 AutoCal DIN
Use two or three different DIN buffer solutions for this proce-
dure in increasing or decreasing order.
Note
Steps 2, 6 and 13 are not required if you use a temperature
probe.
DIN buffer solution
pH
A
1.68
C
4.01
D
6.87
F
9.18
1 Press the <CAL> key repeatedly until
Cd1
and the
AutoCal DIN
function display appear.
2 If necessary, set the temperature of the buffer solu-
tion by pressing <
▲
> <
▼
>
.
3 Submerse the pH electrode in the first buffer solution
.
4 Press the <run/enter> key.
AR
flashes on the display.
The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized,
Cd2
ap-
pears.