3 autocal din – Xylem CG 843 User Manual

Page 32

Advertising
background image

Operation

100

4.3.3 AutoCal DIN

Use two or three different DIN buffer solutions for this proce-
dure in increasing or decreasing order.

Note
Steps 2, 6 and 13 are not required if you use a temperature
probe.

DIN buffer solution

pH

A

1.68

C

4.01

D

6.87

F

9.18

1 Press the <CAL> key repeatedly until

Cd1

and the

AutoCal DIN

function display appear.

2 If necessary, set the temperature of the buffer solu-

tion by pressing <

> <

>

.

3 Submerse the pH electrode in the first buffer solution

.

4 Press the <run/enter> key.

AR

flashes on the display.

The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized,

Cd2

ap-

pears.