Operation

108

4.4.2 Switching on AutoStore (Int 1)

The storage interval (Int 1) determines the time interval
between automatic storage processes.
After the time interval expires, the current data record is
transmitted to the data storage and to the RS interface.

Setting the storage

interval

The storage interval (Int 1) is set to OFF in the factory.
Thus, the

AutoStore

function is switched off.

To switch the function on, set up a time interval (5 s, 10 s,
30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min):

1 Press and hold down the <run/enter> key.

2 Press the <STO> key.

Int.1

appears on the display.

3 Press the <

> <

> keys to set up the required time

interval between storage processes.

4 Confirm with <run/enter>.

The number of free memory locations appears on the
display.