Operation

102

Note
You can also prematurely terminate the three-point calibra-
tion by pressing <pH>. The values of the two-point calibra-
tion for slope and asymmetry are then retained.

11 Press the <run/enter> key.

Cd3

appears on the display.

12 Thoroughly rinse the electrode with distilled water.

13 If necessary, set the temperature of the third buffer

solution by pressing <

> <

>.

14 Submerse the electrode in the third buffer solution.

15 Press the <run/enter> key.

AR

flashes on the display.

The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized,

AR

disap-

pears.
The sensor symbol shows the electrode evaluation
after the three-point calibration.
The value of the slope (mV/pH) appears on the dis-
play.

16 Press the <run/enter> key.

The value of the asymmetry appears on the display
(mV).

17 To return to the measuring mode: Press the <pH>

key.