Xylem CG 843 User Manual

Page 37

Advertising
background image

Operation

105

Single-point

calibration

Use a buffer solution in the range pH = 7.0 ± 0.5 for this pro-
cedure.

Note
Only the electrode asymmetry is determined in single-point
calibration. The slope of the last two-point calibration is
retained.

Note
Step 2 is not required if you use a temperature probe. The
TP message indicates an active temperature measurement.

1 Press the <CAL> key repeatedly until

ASY

and the

ConCal

function display appear.

2 Set the temperature of the buffer solution by pressing

<

> <

>.

3 Submerse the pH electrode in the buffer solution.

4 Press the <run/enter> key.

The measured pH value appears on the display.

5 Set the nominal pH value of the buffer solution (at the

current temperature) by pressing the <

> <

> keys.

6 Press the <run/enter> key. The value of the asym-

metry (mV) and the sensor symbol for the evaluation
of the electrode appears on the display.

7 To return to the measuring mode: Press the <pH>

key.