Temperature probe

Measurements can be performed with and without a temper-
ature probe. A connected temperature probe is indicated by
TP on the display.

Note
The pH meter automatically recognizes the type of the tem-
perature probe used. As a result, you can connect elec-
trodes with the NTC30 or Pt1000.

The temperature measurement is absolutely essential for a
reproducible pH measurement. If the measurement is made
without a temperature probe, proceed as follows:

Note
When calibrating without a temperature probe, set up the
current temperature of the respective buffer solution manu-
ally by pressing the <

> <

> keys.

1 Determine the current temperature using an external

thermometer.

2 Set up the temperature by pressing <

> <

>.