4.4.4

Settings for pH and ORP measurements

Overview

The following settings are possible for pH and ORP measurements:

 Resolution

 Calibration interval

 Buffers for calibration

 Unit of the temperature

 Automatic stability control

 Unit for slope

 Calibration record (display)

Settings

The settings are made in the menu for calibration and measurement
settings of the pH/ORP measurement. To open the settings, display the
required parameter in the measured value display and press the <F1>/
[menu] or <ENTER> key. After completing the settings, switch to the
measured value display with <M>.

Menu item

Possible
setting

Explanation

Calibration / Calibra-
tion record

-

Displays the calibration
record of the last calibration.

Calibration / Buffer

TEC
NIST/DIN
VariCal
...

Buffer sets to be used for pH
calibration.
More buffers and details,
see section 4.4.5.

Calibration / One
point calibration

Yes
No

Quick calibration with 1 buf-
fer

Calibration / Calibra-
tion interval

1 ... 999 d

Calibration interval for the
pH electrode (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring screen.

Calibration / Unit for
slope

mV/pH
%

Unit of the slope.
The % display refers to the
Nernst slope of
-59.2 mV/pH (100 x deter-
mined slope/Nernst slope).

Man. temperature

-25 ... +130 °C

Entry of the manually deter-
mined temperature. For
measurements without tem-
perature sensor only.