What to do if...

HandyLab 100

56

ba77062d01

03/2014

No stable measured

value

Sensor symbol flashes

Display

Cause

Remedy

pH electrode:

– Junction contaminated

– Clean the junction

– Membrane contaminated

– Clean membrane

Test sample:

– pH value not stable

– Measure with air excluded

if necessary

– Temperature not stable

– Adjust temperature if nec-

essary

Electrode + test sample:

– Conductivity too low

– Use suitable electrode

– Temperature too high

– Use suitable electrode

– Organic liquids

– Use suitable electrode

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Batteries almost empty

– Replace the batteries (see

section 5.1 M

AINTENANCE

)