HandyLab 100

Operation

35

ba77062d01

03/2014

Calibration points

Calibration can be performed using one to five buffer solutions in any
order (single-point to five-point calibration). The meter determines the
following values and calculates the calibration line as follows:

Note
You can display the slope in the units, mV/pH or % (see page 30).

Stability control

The calibration procedure automatically activates the stability control
function. The current measurement with stability control can be termi-
nated at any time (accepting the current value).

Calibration record

The new calibration values are displayed when the calibration is fin-
ished.

Display calibration data

You can have the data of the last calibration displayed (see page 44).

Calibration evaluation

After calibrating, the meter automatically evaluates the calibration.The
zero point and slope are evaluated separately. The worse evaluation of
both is taken into account. The evaluation appears on the display and
in the calibration record.

Determined val-
ues

Displayed calibration data

1-point

Asy

 Zero point = Asy
 Slope = Nernst slope

(-59.2 mV/pH at 25 °C)

2-point

Asy
Slp.

 Zero point = Asy
 Slope = Slp.

3-point to
5-point

Asy
Slp.

 Zero point = Asy
 Slope = Slp.

The calibration line is calculated
by linear regression.

Display

Calibration
record

Zero point
[mV]

Slope [mV/
pH]

+++

-15 ... +15

-60,5 ... -58

++

-20 ... +20

-58 ... -57