HandyLab 100
What to do if...
57
ba77062d01
03/2014
Obviously incorrect
measured values
Meter does not react to
keystroke
You want to know which
software
version is in the meter
Cause
Remedy
pH electrode:
– pH electrode unsuitable
– Use suitable electrode
– Temperature difference
between buffer and test sam-
ple too great
– Adjust temperature of buf-
fer or sample solutions
– Measurement procedure not
suitable
– Follow special procedure
Cause
Remedy
– Operating condition unde-
fined or EMC load unallowed
– Processor reset:
Press the <ENTER> and
<ON/OFF> key simultane-
ously
Cause
Remedy
– E.g., a question by the ser-
vice department
– Switch on the meter.
Open the menu, <F1
_
> /
[Menu] / Storage & config /
SystemService informa-
tion. The instrument data
are displayed.