What to do if...

57

ba77062d01

03/2014

Obviously incorrect

measured values

Meter does not react to

keystroke

You want to know which

software

version is in the meter

Cause

Remedy

pH electrode:

– pH electrode unsuitable

– Use suitable electrode

– Temperature difference

between buffer and test sam-
ple too great

– Adjust temperature of buf-

fer or sample solutions

– Measurement procedure not

suitable

– Follow special procedure

Cause

Remedy

– Operating condition unde-

fined or EMC load unallowed

– Processor reset:

Press the <ENTER> and
<ON/OFF> key simultane-
ously

Cause

Remedy

– E.g., a question by the ser-

vice department

– Switch on the meter.

Open the menu, <F1

_

> /

[Menu] / Storage & config /
SystemService informa-
tion
. The instrument data
are displayed.