Operation

HandyLab 100

44

ba77062d01

03/2014

Note
After the fifth buffer has been measured the calibration is automatically
finished. Then the calibration record is displayed.

The calibration line is determined by linear regression.

4.4.9

Displaying calibration records

Displaying the

calibration record

The calibration record of the last calibration is to be found under the
menu item, Calibration / Calibration record. To open it in the measured
value display, press the <CAL

_

> key.

22

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.

23

Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <ENTER>.
The calibration display for the next buffer appears.

24

Set the nominal buffer value for the measured temperature with
<

><>.

25

Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).

26

Use <M> to finish calibration or
switch to calibration with the next buffer with <ENTER>.

pH

9.958

24.8 °C

VariCal

3

15.03.2014

010:00

Buffer