6 calibration interval, Calibration interval – Xylem HandyLab 100 User Manual

Page 36

Advertising
background image

Operation

HandyLab 100

36

ba77062d01

03/2014

Preparatory activities

Perform the following preparatory activities when you want to calibrate:

4.4.6

Calibration interval

The calibration evaluation is displayed as a sensor symbol.

The sensor symbol flashes after the adjusted calibration interval has
expired. It is still possible to measure.

Note
To ensure the high measuring accuracy of the measuring system, cal-
ibrate after the calibration interval has expired.

Setting the calibration

interval

The calibration interval is set to 7 days (d7) in the factory.
You can change the interval (1 ... 999 days):

+

-25 ... +25

-61 ... -60.5
or
-57 ... -56

-

-30 ... +30

-62 ... -61
or
-56 ... -50

Clean the electrode according to the
electrode operating manual

Error

Error

< -30 or
> 30

... -62 or
... -50

Eliminate the error according to chapter
6 W

HAT

TO

DO

IF

... (page 55)

Display

Calibration
record

Zero point
[mV]

Slope [mV/
pH]

1

Connect the pH electrode to the meter.
The pH measuring window is displayed.

2

Keep the buffer solutions ready. Adjust the temperature of the
buffer solutions, or measure the current temperature, if you
measure without a temperature sensor.

1

Open the menu for measurement settings with [Menu]<F1>.

2

In the Calibration / Calibration interval menu, set the calibration
interval with <

><>.