2 measuring ranges, resolution, accuracy, Measuring ranges, resolution, accuracy – Xylem HandyLab 100 User Manual

Page 60

Advertising
background image

Technical data

HandyLab 100

60

ba77062d01

03/2014

7.2

Measuring ranges, resolution, accuracy

Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the electrodes and buffer solutions has to be taken into
account additionally.

Measuring ranges,

resolution

Variable

Measuring range

Resolution

pH

- 2.0 ... + 20.0

0.1

- 2.00 ... + 20.00

0.01

- 2.000 ... + 19.999

0.001

U [mV]

- 2500 ... + 2500

1

- 1200.0 ... + 1200.0

0.1

T [°C]

- 5.0 ... + 105.0

0.1

T [°F]

23.0 ... + 221.0

0.1

Manual

temperature input

Variable

Range

Increment

T

manual

[°C]

- 25 ... + 130

1

T

manual

[°F]

-13 ... + 266

1

Accuracy (± 1 digit)

Variable

Accuracy

Temperature of the test
sample

pH / range *

- 2.0 ... + 20.0

± 0.1

+ 15 °C ... + 35 °C

- 2.00 ... + 20.00

± 0.01

+ 15 °C ... + 35 °C

- 2.000 ... + 19.999

± 0.005

+ 15 °C ... + 35 °C

U [mV] / range

- 2500 ... + 2500

± 1

+ 15 °C ... + 35 °C

-1200.0 ... +1200.

± 0.3

+ 15 °C ... + 35 °C

T [°C] / temperature sensor

NTC 30

± 0.1

PT 1000

± 0.1

* when measuring in a range of ± 2 pH around a calibration point