HandyLab 100

Operation

43

ba77062d01

03/2014

Continuing with two-

point calibration

Continuing with three-

to five-point calibration

11

Thoroughly rinse the electrode with deionized water.

12

Immerse the electrode in buffer solution 2.

13

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <

><>.

14

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.

15

Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <ENTER>.
The calibration display for the next buffer appears.

16

Set the nominal buffer value for the measured temperature with
<

><>.

17

Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).

18

Finish the calibration procedure as a two-point calibration with
<M>.
The calibration record is displayed.

pH

4.035

24.8 °C

VariCal

2

15.03.2014

010:00

Buffer


19

Thoroughly rinse the electrode with deionized water.

20

Immerse the electrode in the next buffer solution.

21

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <

><>.