6 reset, 1 resetting the measurement settings, Reset 4.6.1 – Xylem HandyLab 100 User Manual
Page 49: Resetting the measurement settings

HandyLab 100
Operation
49
ba77062d01
03/2014
4.6
Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.6.1
Resetting the measurement settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
pH
The following settings for pH measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under the Reset menu item in the menu
for calibration and measurement settings. To open the settings, display
the required parameter in the measured value display and press the
<F1>/[menu] or <ENTER> key.
Setting
Default settings
Buffer TEC
Cal. interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
Resolution pH
0.001
Resolution mV
0.1
Asymmetry
0 mV
Slope
-59.2 mV
Man. temperature
25 °C
One point calibration
Off
Stability control
On
Temperature unit
°C