6 reset, 1 resetting the measurement settings, Reset 4.6.1 – Xylem HandyLab 100 User Manual

Page 49: Resetting the measurement settings

Advertising
background image

HandyLab 100

Operation

49

ba77062d01

03/2014

4.6

Reset

You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.

4.6.1

Resetting the measurement settings

Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.

pH

The following settings for pH measurements are reset to the default
settings with the Reset function:

The sensor settings are reset under the Reset menu item in the menu
for calibration and measurement settings. To open the settings, display
the required parameter in the measured value display and press the
<F1>/[menu] or <ENTER> key.

Setting

Default settings

Buffer TEC

Cal. interval

7 d

Unit for slope

mV/pH

Measured parameter

pH

Resolution pH

0.001

Resolution mV

0.1

Asymmetry

0 mV

Slope

-59.2 mV

Man. temperature

25 °C

One point calibration

Off

Stability control

On

Temperature unit

°C